Fictional Nursery Alert Timing

Fictional Nursery Alert Timing

Try Press Step to move from baseline to the warning trend and watch whether either alert clock starts as the reading falls.

Fictional Nursery Alert Timing

Step-through

Given Maya’s fictional trace, decide when an averaged SpO₂ reading can qualify for an urgent alert, identify when the phone-delivery clock starts, and explain what evidence is required for resolution.

Fictional device workflow | Not infant triage advice.

Upstream processing

4-second averaging

Chapter SpO₂ trace

Walkthrough assumption: the values shown are processed inputs to the alert rule.

  1. 0–15 s97–98%
  2. 16–24 s90–94%
  3. 25–34 s78–79%
  4. 35–45 s75–77%
  5. 46–60 s82–95%

Hub · Urgent rule

Persistence timer idle

Phone delivery clock

Idle · starts at urgent decision

At baseline, the rule evaluates the processed reading. Averaging describes how the readings are processed. Persistence and delivery each have their own clock.

Observe At 78–79%, the hub shows 10 s elapsed with the urgent rule pending. In the 75–77% band, the urgent alert appears and the phone delivery clock starts.
Explain The urgent rule requires continuous valid processed readings below 80% for more than 10 s, so the first eligible decision after that boundary starts a separate clock for phone delivery.

Technical boundaries. This fictional workflow uses processed SpO₂ values and assumes continuous valid input through the low bands. The phone card gives a delivery target and leaves the outcome unverified. Recovery also depends on a configured window and a recorded caregiver response.

Back to the chapter