Polynomial Calibration Explorer

Polynomial Calibration Explorer

animation
sensors
Try Leave the five reference pairs in place and move Test Raw Reading (°C) from 40.0. Then change Raw 3 (°C) from 52.1 and compare the fit and residual table.
Observe Test Raw Reading (°C) starts at 40.0, with its corrected value shown below the fitted formula. Moving the test reading changes that result; editing a calibration pair also changes the coefficients and residuals.
Explain The five pairs determine a least-squares quadratic from raw to reference values; Test Raw Reading (°C) is substituted into that polynomial to produce the corrected reading.

Technical boundaries. The fit assumes all five pairs carry equal weight and always uses a quadratic, with controls limited to 0-100 degrees C. A determinant threshold rejects a singular fit; the residuals come from the fitting points rather than independent validation data.

Polynomial Calibration Explorer

When a straight line leaves curved errors, a richer model may help. The size of a coefficient alone does not establish whether the response is linear. Small errors at the points used for a fit do not prove accuracy elsewhere. Test the result at separate reference points.

Note

Enter 3–5 calibration points (reference vs raw) and see how a quadratic polynomial fits the data. Adjust individual points to observe how nonlinearity affects the correction curve.

Back to the chapter