TryLeave Calibration scale factor at 1.0817 and Calibration offset at -2.38, then compare Application tolerance at +/-0.5°C (HVAC/Greenhouse) and +/-2.0°C (Rough estimate).
ObserveThe opening 21.45°C raw reading becomes 20.82°C against a 22.5°C reference. Its 1.678°C error fails the default tolerance but passes the wider one.
ExplainCalibration scale factor multiplies the raw reading before Calibration offset is added; the absolute reference difference must be no greater than the chosen tolerance.
Technical boundaries. Validation assumes a linear correction and a single independent reference point. The coefficients are supplied by the user, and no uncertainty is assigned to the reference or sensor reading.
Try It: Calibration Validation Checker
A separate test point challenges the fitted correction. Use the checker to compare its error with a limit.
Note
After calibrating a sensor, test it against a third reference point to verify accuracy. Enter your calibration coefficients and a validation measurement to determine if your calibration is adequate or if you need a different approach.
viewof val_scale = Inputs.range([0.5,1.5], {value:1.0817,step:0.0001,label:"Calibration scale factor"})viewof val_offset = Inputs.range([-10,10], {value:-2.38,step:0.01,label:"Calibration offset"})viewof val_raw_reading = Inputs.range([-20,120], {value:21.45,step:0.05,label:"Raw sensor reading at validation point (°C)"})viewof val_true_value = Inputs.range([-20,120], {value:22.5,step:0.1,label:"True reference value at validation point (°C)"})viewof val_app_tolerance = Inputs.select( ["+/-0.2°C (Laboratory)","+/-0.5°C (HVAC/Greenhouse)","+/-1.0°C (General monitoring)","+/-2.0°C (Rough estimate)"], {value:"+/-0.5°C (HVAC/Greenhouse)",label:"Application tolerance"})