Calibration Reference Checkpoints
Calibration Reference Checkpoints
animation
sensors
Tip
Checkpoint: Error Shape First
Checkpoint: Error Shape First
You now know:
- Offset, gain, nonlinearity, noise, and drift are different failure modes, so they should not all receive the same correction.
- A 0.5°C disagreement between two DHT22 sensors calls for comparison with a trusted reference; it does not identify which sensor is wrong.
- Calibration starts with a trusted reference and ends with evidence that the correction still holds.
Tip
Checkpoint: Linear Calibration Math
Checkpoint: Linear Calibration Math
You now know:
- One-point calibration corrects a constant offset such as 2.3°C.
- Two-point calibration uses two references, such as 0°C and 100°C, to compute both scale and offset.
- In the worked example, raw readings of 1.8°C and 98.5°C produce scale
1.034and offset-1.86.