Calibration Reference Checkpoints

Calibration Reference Checkpoints

animation
sensors
TipPhysics Phoebe, Physics and measurement-foundations expertCheckpoint: Error Shape First

You now know:

  • Offset, gain, nonlinearity, noise, and drift are different failure modes, so they should not all receive the same correction.
  • A 0.5°C disagreement between two DHT22 sensors calls for comparison with a trusted reference; it does not identify which sensor is wrong.
  • Calibration starts with a trusted reference and ends with evidence that the correction still holds.
TipPhysics Phoebe, Physics and measurement-foundations expertCheckpoint: Linear Calibration Math

You now know:

  • One-point calibration corrects a constant offset such as 2.3°C.
  • Two-point calibration uses two references, such as 0°C and 100°C, to compute both scale and offset.
  • In the worked example, raw readings of 1.8°C and 98.5°C produce scale 1.034 and offset -1.86.

Back to the chapter